Application of computer-aided diagnosis for Lung-RADS categorization in CT screening for lung cancer: effect on inter-reader agreement
- Abstract
- Objectives: To evaluate the effects of computer-aided diagnosis (CAD) on inter-reader agreement in Lung Imaging Reporting and Data System (Lung-RADS) categorization.
Methods: Two hundred baseline CT scans covering all Lung-RADS categories were randomly selected from the National Lung Cancer Screening Trial. Five radiologists independently reviewed the CT scans and assigned Lung-RADS categories without CAD and with CAD. The CAD system presented up to five of the most risk-dominant nodules with measurements and predicted Lung-RADS category. Inter-reader agreement was analyzed using multirater Fleiss κ statistics.
Results: The five readers reported 139-151 negative screening results without CAD and 126-142 with CAD. With CAD, readers tended to upstage (average, 12.3%) rather than downstage Lung-RADS category (average, 4.4%). Inter-reader agreement of five readers for Lung-RADS categorization was moderate (Fleiss kappa, 0.60 [95% confidence interval, 0.57, 0.63]) without CAD, and slightly improved to substantial (Fleiss kappa, 0.65 [95% CI, 0.63, 0.68]) with CAD. The major cause for disagreement was assignment of different risk-dominant nodules in the reading sessions without and with CAD (54.2% [201/371] vs. 63.6% [232/365]). The proportion of disagreement in nodule size measurement was reduced from 5.1% (102/2000) to 3.1% (62/2000) with the use of CAD (p < 0.001). In 31 cancer-positive cases, substantial management discrepancies (category 1/2 vs. 4A/B) between reader pairs decreased with application of CAD (pooled sensitivity, 85.2% vs. 91.6%; p = 0.004).
Conclusions: Application of CAD demonstrated a minor improvement in inter-reader agreement of Lung-RADS category, while showing the potential to reduce measurement variability and substantial management change in cancer-positive cases.
- Author(s)
- Sohee Park; Hyunho Park; Sang Min Lee; Yura Ahn; Wooil Kim; Kyuhwan Jung; Joon Beom Seo
- Issued Date
- 2022
- Type
- Article
- Keyword
- Computer-assisted; Diagnostic screening program; Lung neoplasms; Observer variation; Tomography; X-ray computed
- DOI
- 10.1007/s00330-021-08202-3
- URI
- https://oak.ulsan.ac.kr/handle/2021.oak/14108
- Publisher
- EUROPEAN RADIOLOGY
- Language
- 영어
- ISSN
- 0938-7994
- Citation Volume
- 32
- Citation Number
- 2
- Citation Start Page
- 1054
- Citation End Page
- 1064
-