Objectives: Complications like complete atrioventricular block (CAVB) and paravalvular leakage (PVL) following rapid deployment aortic valve (AV) replacement (RDAVR) remain unresolved. Selecting an optimal size of the valve might be important to minimize the incidence of these complications. We sought to determine the impact of prosthesis size relative to the anatomic profile of the AV on the occurrence of CAVB or PVL after RDAVR.

Methods: Preoperative computed tomographic (CT) images were evaluated in patients receiving RDAVR (INTUITY ELITE) between February 2016 and December 2019. The occurrence of CAVB requiring permanent pacemaker implantation and PVL (≥ mild) was evaluated. The relative size of implants against the cross-sectional dimensions of recipients' AV annulus and left ventricular outflow tract (LVOT) were calculated.

Results: Among 187 eligible patients, CAVB and PVL (≥ mild) occurred in 12 (6.4%) and 11 patients (5.9%), respectively. CAVB was associated with oversized RDAVR (RDAVR frame width minus average diameter of LVOT calculated from the cross-sectional area [ΔLVOTarea]: odds ratio, 2.05; 95% confidence interval, 1.28-3.30): this was with an area under the curve of 0.78 (P = 0.005). The projected probability of CAVB was <3% when the ΔLVOTarea was <1.3. In contrast, PVL was associated with under-sized RDAVR (RDAVR size divided by the longest diameter of AV annulus [index Annlong]: odds ratio, 0.64; 95% confidence interval, 0.51-0.79): This was with an area under curve of 0.94 (P < 0.001).

Conclusions: CT parameters of the AV annulus and LVOT are highly reliable in the prediction of CAVB or PVL after RDAVR. Our data might justify CT-based sizing of prosthesis for RDAVR.
Author(s)
Sung Jun ParkYounju RheeChee-Hoon LeeHo Jin KimJoon Bum KimSuk Jung ChooJae Won Lee
Issued Date
2022
Type
Article
Keyword
Complete atrio-ventricular blockComputed tomographyLeft ventricular outflow tractParavalvular leakageRapid deployment aortic valve replacement
DOI
10.1093/ejcts/ezab431
URI
https://oak.ulsan.ac.kr/handle/2021.oak/14165
Publisher
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY
Language
영어
ISSN
1010-7940
Citation Volume
61
Citation Number
4
Citation Start Page
899
Citation End Page
907
Appears in Collections:
Medicine > Nursing
공개 및 라이선스
  • 공개 구분공개
파일 목록
  • 관련 파일이 존재하지 않습니다.

Items in Repository are protected by copyright, with all rights reserved, unless otherwise indicated.