Effect of Outlier Elimination on the 99th Percentile Upper Reference Limits of High-Sensitivity Cardiac Troponin I Assays Based on a Strictly Selected Healthy Reference Population
- Abstract
- Background: We established high-sensitivity cardiac troponin I (hsTnI) 99th percentile upper reference limits (URLs) for the Centaur XPT High-Sensitivity Troponin I assay (Centaur hsTnI; Siemens, Erlangen, Germany) and Atellica IM High-Sensitivity Troponin I assay (Atellica hsTnI; Siemens) and assessed the effect of outlier elimination.
Methods: The reference population comprised 380 men and 387 women, satisfying the strict systematic reference population criteria. After reference population verification by the N-terminal pro-B-type natriuretic peptide (NT-proBNP) assay, 99th percentile URLs for Centaur hsTnI and Atellica hsTnI were calculated before and after outlier elimination.
Results: The 99th percentile URL for Centaur hsTnI was 60.4 (men, 74.7; women, 57.5) ng/L and that for Atellica hsTnI was 59.6 (men, 75.2; women, 55.1) ng/L. After the elimination of 61 (8.0%) outlier samples in Centaur hsTnI and 58 (7.6%) in Atellica hsTnI, the 99th percentile URLs were 13.5 ng/L (men, 15.3 ng/L; women, 11.9 ng/L) and 13.4 ng/L (men, 15.5 ng/L; women, 12.9 ng/L), respectively, significantly lower than those before outlier elimination. The CVs at the 99th percentile URLs were 5.2% and 3.5%, respectively. The measurable fractions among the reference population were 91.5% and 93.4%, respectively. Performance evaluation of Atellica B-type natriuretic peptide (BNP), Atellica NT-proBNP, Centaur hsTnI, and Atellica hsTnI showed outstanding results.
Conclusions: The Korean hsTnI 99th percentile URLs calculated in this study were significantly lower after outlier elimination than before. Centaur hsTnI and Atellica hsTnI meet the "Guideline acceptable" and "Level 3 (second generation, high sensitivity)" requirements, satisfying international standards.
- Author(s)
- Sunyoung Ahn; Hyun-Ki Kim; Woochang Lee; Sail Chun; Won-Ki Min
- Issued Date
- 2022
- Type
- Article
- Keyword
- Ninety-ninth percentile; Upper reference limit; Korean; Outlier; ADVIA Centaur XPT High-Sensitivity Troponin I assay; Atellica IM High-Sensitivity Troponin I assay; Key Words
- DOI
- 10.3343/alm.2022.42.3.331
- URI
- https://oak.ulsan.ac.kr/handle/2021.oak/15489
- Publisher
- ANNALS OF LABORATORY MEDICINE
- Language
- 한국어
- ISSN
- 2234-3806
- Citation Volume
- 42
- Citation Number
- 3
- Citation Start Page
- 331
- Citation End Page
- +
-
Appears in Collections:
- Medicine > Nursing
- 공개 및 라이선스
-
- 파일 목록
-
Items in Repository are protected by copyright, with all rights reserved, unless otherwise indicated.