Background: Drug-coated balloons (DCBs) offer an effective treatment for in-stent restenosis (ISR). The Genoss DCB is a novel paclitaxel-coated balloon with a shellac plus vitamin E excipient that enhances drug delivery to the target lesion, minimizing restenosis.

Objectives: This study aimed to investigate the angiographic efficacy, clinical safety, and effectiveness of the novel shellac plus vitamin E-based DCB in a randomized controlled trial designed to enable regulatory approval of this new device in South Korea.

Methods: This noninferiority trial randomized patients experiencing their first ISR to the novel shellac plus vitamin E-based DCB or the reference SeQuent Please iopromide-based DCB in a 1:1 ratio. All patients underwent planned angiographic and clinical follow-up at 6 months. The study was powered for the primary endpoint of 6 months in-segment late lumen loss (LLL).

Results: A total of 82 patients from 7 centers were randomized to either the novel shellac plus vitamin E-based DCB group (n = 41) or the reference iopromide-based DCB group (n = 41). The 6-month in-segment LLL was 0.15 ± 0.43 mm with the novel DCB compared with 0.24 ± 0.39 mm with the reference device. The 1-sided 97.5% upper confidence limit of the difference was 0.13 mm, lower than the noninferiority limit of 0.29 mm, achieving noninferiority (P for noninferiority = 0.001). Major cardiovascular events were comparable between 2 groups at 6 months (7.7% for the novel DCB vs 10.3% for the reference DCB; P = 0.692).

Conclusions: In this multicenter, head-to-head comparison randomized trial, the novel shellac plus vitamin E-based DCB showed a comparable result to the reference iopromide-based device for the primary endpoint of 6-month in-segment LLL for the treatment of coronary ISR.
Author(s)
Eun Jung JunEun-Seok ShinSong Lin YuanYoungjune BhakScot GargWoong Chol KangJe Sang KimJune-Hong KimJang-Whan BaeSeung-Woon RhaIn-Ho Chae
Issued Date
2022
Type
Article
Keyword
DCBdrug-coated balloonDESdrug-eluting stentDSdiameter stenosisISRin-stent restenosisLLLlate lumen lossMACEmajor adverse cardiovascular eventMLDminimal lumen diameterPCBpaclitaxel-coated balloonTLRtarget lesion revascularizationcardiovascular eventsin-segment late lumen lossmajor adverse
DOI
10.1016/j.jacasi.2021.11.015
URI
https://oak.ulsan.ac.kr/handle/2021.oak/15737
Publisher
JACC: Asia
Language
영어
ISSN
2772-3747
Citation Volume
2
Citation Number
2
Citation Start Page
170
Citation End Page
179
Appears in Collections:
Medicine > Nursing
공개 및 라이선스
  • 공개 구분공개
파일 목록
  • 관련 파일이 존재하지 않습니다.

Items in Repository are protected by copyright, with all rights reserved, unless otherwise indicated.