Objectives
This study assessed the clinical performance of the cobas Liat SARS‑CoV‑2 & Influenza A/B assay (LiatCOVID/flu) for the detection of both severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) and influenza viruses during the SARS-CoV-2 Omicron outbreak.

Methods
Residual nasopharyngeal swab samples (NPS) previously tested with cobas SARS-CoV-2 & Influenza A/B for SARS-CoV-2 and with the Allplex Respiratory Panel 1 for influenza viruses were collected. All samples were submitted to the LiatCOVID/flu assay.

Results
A total of 1147 samples were collected comprising 167 SARS-CoV-2-positive, 556 SARS-CoV-2-negative, 224 influenza-positive, and 200 influenza-negative cases. The positive percent agreement (PPA)/negative percent agreement (NPA) of LiatCOVID/flu for SARS-CoV-2 and influenza viruses compared to the previously tested methods were 100% of 100% and 99.6% of 100%, respectively.

Conclusions
The LiatCOVID/flu assay shows an acceptable performance in the detection of SARS-CoV-2 and influenza viruses using NPS samples.
Issued Date
2023
Kuenyoul Park
Heungsup Sung
Mi-Na Kim
Type
Article
Keyword
SARS-CoV-2COVID-19POCTrapid diagnostic testinfluenzanasopharyngeal swab
DOI
10.1016/j.diagmicrobio.2023.115891
URI
https://oak.ulsan.ac.kr/handle/2021.oak/15940
Publisher
DIAGNOSTIC MICROBIOLOGY AND INFECTIOUS DISEASE
Language
한국어
ISSN
0732-8893
Citation Volume
105
Citation Number
4
Citation Start Page
1
Citation End Page
5
Appears in Collections:
Medicine > Nursing
공개 및 라이선스
  • 공개 구분공개
파일 목록
  • 관련 파일이 존재하지 않습니다.

Items in Repository are protected by copyright, with all rights reserved, unless otherwise indicated.