Dosimetric verification of annual quality assurance for a linear accelerator using a transmission type detector
- Abstract
- The purpose of our study is to establish an efficient quality assurance (QA) procedure using a transmission-type detector (IBA, Stealth chamber), a reference signal detector, as a field chamber. Relative dosimetry items, including monitor unit linearity, output constancy based on dose rate and field size, and output factor were measured and compared with results obtained from the Farmer-type chamber (IBA, Wellhofer, FC65-G). Moreover, output for each field size was measured to assess its applicability to small fields. Results using the Stealth chamber were in good agreement with the FC65-G within 1.0%, except for output constancy according to gantry angle, which had a 1.1% error rate for the Stealth chamber and 2.7% for the FC65-G. Differences of up to − 6.26% output factor were observed for the Stealth chamber and up to − 0.56% for the CC-13 ionization chamber (IBA) in the 3 × 3 cm2 field. Our study confirmed the possibility of using Stealth chambers for relative dosimetry measurement in QA.
- Issued Date
- 2023
Dong Hyeok Choi
Jin Sung Kim
Rena Lee
So Hyun Ahn
Woo Sang Ahn
- Type
- Article
- Keyword
- Ionization; Quality assurance; Quality control; Science
- DOI
- 10.1038/s41598-023-45114-2
- URI
- https://oak.ulsan.ac.kr/handle/2021.oak/16999
- Publisher
- SCIENTIFIC REPORTS
- Language
- 영어
- ISSN
- 2045-2322
- Citation Volume
- 13
- Citation Number
- 1
- Citation Start Page
- 1
- Citation End Page
- 13
-
Appears in Collections:
- Medicine > Nursing
- 공개 및 라이선스
-
- 파일 목록
-
Items in Repository are protected by copyright, with all rights reserved, unless otherwise indicated.