새로 개발된 Multi-leaf Collimator의 방사선 차폐효과
- Alternative Title
- The radiation shielding characteristics of the newly developed type Multi-leaf Collimator
- Abstract
- We have newly developed Multi-leaf Collimator(MLC), which is the one of the essential equipments for coformal therapy, and examined its radiation shielding characteristics.
Radiation transmission through leaf is 1.5∼1.9% for various photon energies(4MV, 6MV and 15MV) at the dmax depth. And those for various depths are around 2% : The transmission factor is increased as photon energy increases, while one cannot find depth-transmission relation. Inter-leaf leakage is 1.9∼2.5%, which is good for clinical application.
AMC type MLC shows superior shielding characteristics to commercially available MLC system.
We have newly developed Multi-leaf Collimator(MLC), which is the one of the essential equipments for coformal therapy, and examined its radiation shielding characteristics.
Radiation transmission through leaf is 1.5∼1.9% for various photon energies(4MV, 6MV and 15MV) at the dmax depth. And those for various depths are around 2% : The transmission factor is increased as photon energy increases, while one cannot find depth-transmission relation. Inter-leaf leakage is 1.9∼2.5%, which is good for clinical application.
AMC type MLC shows superior shielding characteristics to commercially available MLC system.
- Author(s)
- 이병용; 최은경; 김종훈; 장혜숙
- Issued Date
- 1995
- Type
- Research Laboratory
- URI
- https://oak.ulsan.ac.kr/handle/2021.oak/5259
http://ulsan.dcollection.net/jsp/common/DcLoOrgPer.jsp?sItemId=000002024285
- 공개 및 라이선스
-
- 파일 목록
-
Items in Repository are protected by copyright, with all rights reserved, unless otherwise indicated.