KLI

전자선의 Virtual Source Distance의 공기중 측정법

Metadata Downloads
Alternative Title
The Air Scanning of Electron Beams for Determination of the Virtual Source Distance
Abstract
Authors have measured virtual source distance of electron beam from CL/1800 medical linear accelerator, with newly designed method. Beam scanning was performed with the direction of beam axis in the air. Compared results between this study and well established in-phamtom measurement showed good agreement within experiment error.

And we have found that build-up cap plays very important role in air measurement because of charge build up.

The nethod of in-air measurement of virtual source distance is very easy to setup and generate accurate results.
Authors have measured virtual source distance of electron beam from CL/1800 medical linear accelerator, with newly designed method. Beam scanning was performed with the direction of beam axis in the air. Compared results between this study and well established in-phamtom measurement showed good agreement within experiment error.

And we have found that build-up cap plays very important role in air measurement because of charge build up.

The nethod of in-air measurement of virtual source distance is very easy to setup and generate accurate results.
Author(s)
양광모이병용최은경장혜숙
Issued Date
1993
Type
Research Laboratory
URI
https://oak.ulsan.ac.kr/handle/2021.oak/5296
http://ulsan.dcollection.net/jsp/common/DcLoOrgPer.jsp?sItemId=000002024718
Alternative Author(s)
Yang,Kwang MoYi,Byong YongChoi,Eun KyungChang,Hye sook
Publisher
울산의대학술지
Language
kor
Rights
울산대학교 저작물은 저작권에 의해 보호받습니다.
Citation Volume
2
Citation Number
1
Citation Start Page
79
Citation End Page
82
Appears in Collections:
Research Laboratory > The ULSAN university medical journal
공개 및 라이선스
  • 공개 구분공개
파일 목록

Items in Repository are protected by copyright, with all rights reserved, unless otherwise indicated.