Value of bone suppression software in chest radiographs for improving image quality and reducing radiation dose

Metadata Downloads
Abstract
Objectives To compare image quality and radiation dose between dual-energy subtraction (DES)-based bone suppression images (D-BSIs) and software-based bone suppression images (S-BSIs).

Methods Chest radiographs (CXRs) of forty adult patients were obtained with the two X-ray devices, one with DES and one with bone suppression software. Three image quality metrics (relative mean absolute error (RMAE), peak signal-to-noise ratio (PSNR), and structural similarity index (SSIM)) between original CXR and BSI for each of D-BSI and S-SBI groups were calculated for each bone and soft tissue areas. Two readers rated the visual image quality for original CXR and BSI for each of D-BSI and S-SBI groups. The dose area product (DAP) values were recorded. Paired t test was used to compare the image quality and DAP values between D-BSI and S-BSI groups.

Results In bone areas, S-BSIs had better SSIM values than D-BSI (94.57 vs. 87.77) but worse RMAE and PSNR values (0.50 vs. 0.20; 20.93 vs. 34.37) (all p < 0.001). In soft tissue areas, S-BSIs had better SSIM values than D-BSI (97.56 vs. 91.42) but similar RMAE and PSNR values (0.29 vs. 0.27; 31.35 vs. 29.87) (all p < 0.001). Both readers gave S-BSIs significantly higher image quality scores than D-BSI (p < 0.001). The mean DAP in software-related images (0.98 dGy.cm(2)) was significantly lower than that in the DES-related images (1.48 dGy.cm(2)) (p < 0.001).

Conclusion Bone suppression software significantly improved the image quality of bone suppression images with a relatively lower radiation dose, compared with dual-energy subtraction technique.
Author(s)
도경현이충욱조경욱홍길선A-Yeon SonJihye YunKwang Pyo Kim
Issued Date
2021
Type
Article
Keyword
ChestDiagnostic RadiologyImage processingImagingInternal MedicineInterventional RadiologyMedicineMedicine &amp; Pubic HealthNeuroradiologyRadiationRadiologyUltrasound
DOI
10.1007/s00330-020-07596-w
URI
https://oak.ulsan.ac.kr/handle/2021.oak/7157
https://ulsan-primo.hosted.exlibrisgroup.com/primo-explore/fulldisplay?docid=TN_cdi_proquest_miscellaneous_2477512493&amp;context=PC&amp;vid=ULSAN&amp;lang=ko_KR&amp;search_scope=default_scope&amp;adaptor=primo_central_multiple_fe&amp;tab=default_tab&amp;query=any,contains,Value%20of%20bone%20suppression%20software%20in%20chest%20radiographs%20for%20improving%20image%20quality%20and%20reducing%20radiation%20dose&amp;offset=0&amp;pcAvailability=true
Publisher
EUROPEAN RADIOLOGY
Location
독일
Language
영어
ISSN
0938-7994
Citation Volume
31
Citation Number
7
Citation Start Page
5160
Citation End Page
5171
Appears in Collections:
Medicine > Medicine
공개 및 라이선스
  • 공개 구분공개
파일 목록
  • 관련 파일이 존재하지 않습니다.

Items in Repository are protected by copyright, with all rights reserved, unless otherwise indicated.