Reproducibility of radiomic features in SENSE and compressed SENSE: impact of acceleration factors

Metadata Downloads
Abstract
Objectives To investigate the impact of acceleration factors on reproducibility of radiomic features in sensitivity encoding (SENSE) and compressed SENSE (CS), compare between SENSE and CS, and identify reproducible radiomic features. Methods Three-dimensional turbo spin echo T1-weighted imaging was performed in 14 healthy volunteers (mean age, 57 years; range, 33-67 years; 7 men) under SENSE and CS with accelerator factors of 5.5, 6.8, and 9.7. Eight anatomical locations (brain parenchyma, salivary glands, masseter muscle, tongue, pharyngeal mucosal space, eyeballs) were evaluated. Reproducibility of radiomic features was evaluated by calculating concordance correlation coefficient (CCC) in reference to the original image (SENSE with acceleration factor of 3.5). Reproducibility of radiomic features among acceleration factors and between SENSE and CS was compared. Results Proportion of radiomic features with CCC > 0.85 in reference to the original image was lower with higher acceleration factors in both SENSE and CS across all anatomical locations (p < .001). Proportion of radiomic features with CCC > 0.85 in reference to the original image was higher in SENSE compared with CS (SENSE, 6.7-7.3% vs CS, 4.4-5.0%; p < .001). Run percentage of gray-level run-length matrix (GLRLM) with wavelet D showed CCC > 0.85 in reference to the original image in both SENSE and CS at acceleration factor of 9.7 in the highest number of anatomical locations. Conclusions Higher acceleration factors resulted in lower reproducibility of radiomic features in both SENSE and CS, and SENSE showed higher reproducibility of radiomic features than CS in reference to the original image. Run percentage of GLRLM with wavelet D was identified as the most reproducible feature.
Author(s)
김민재박서영박지은이현나정승채최금미
Issued Date
2021
Type
Article
Keyword
Diagnostic RadiologyHead and NeckImagingInternal MedicineInterventional RadiologyMedicineMedicine &amp; Public HealthNeuroradiologyRadiologyUltrasound
DOI
10.1007/s00330-021-07760-w
URI
https://oak.ulsan.ac.kr/handle/2021.oak/7430
https://ulsan-primo.hosted.exlibrisgroup.com/primo-explore/fulldisplay?docid=TN_cdi_proquest_miscellaneous_2502806805&amp;context=PC&amp;vid=ULSAN&amp;lang=ko_KR&amp;search_scope=default_scope&amp;adaptor=primo_central_multiple_fe&amp;tab=default_tab&amp;query=any,contains,Reproducibility%20of%20radiomic%20features%20in%20SENSE%20and%20compressed%20SENSE:%20impact%20of%20acceleration%20factors&amp;offset=0&amp;pcAvailability=true
Publisher
EUROPEAN RADIOLOGY
Location
독일
Language
한국어
ISSN
0938-7994
Citation Volume
31
Citation Number
9
Citation Start Page
6457
Citation End Page
6470
Appears in Collections:
Medicine > Medicine
공개 및 라이선스
  • 공개 구분공개
파일 목록
  • 관련 파일이 존재하지 않습니다.

Items in Repository are protected by copyright, with all rights reserved, unless otherwise indicated.