Dosimetric Comparison between CyberKnife and GammaKnife Hypofractionated Radiosurgery for Benign Perioptic Tumors: A Retrospective Multi-institutional Study
- Alternative Title
- Dosimetric Comparison between CyberKnife and GammaKnife Hypofractionated Radiosurgery for Benign Perioptic Tumors: A Retrospective Multi-institutional Study
- Abstract
- Objective: We conducted a retrospective multi-institutional study to compare the dosimetric differences between the CyberKnife (CK) and the Gamma Knife (GK) in treatment planning of hypofractionated stereotactic radiosurgery (hSRS) for benign perioptic tumors.
Methods: We selected 16 perioptic tumors previously treated using CK between 2011 and 2015. In each case, two separate GK plans (GK1 and GK2) at two different institutions were retrospectively generated using the identical planning CT images and contour set of the original CK plan. Thus, a triplet data set of dosimetric parameters for each lesion was generated and compared. The planning objective was maximizing dose to tumor, while satisfying the proposed dose-volume constraints criteria (Dmax < 25 Gy and V20Gy < 0.2 cc) for the optic apparatus (OA).
Results: All treatment plans achieved a high level of conformity index (1.22 for CK, 1.21 for GK1, and 1.20 for GK2 plan, respectively; P=0.025), while doses to OA were well within the intended dose-volume limits. However, statistically significant differences in other dosimetric parameters were observed: D99%, 25.6 Gy for CK vs. 24.9 Gy and 23.9 Gy for GK1 and GK2, respectively (p < 0.001); tumor coverage, 95.1% for CK vs. 90.5% and 95.1% for GK1 and GK2, respectively (p = 0.002); and gradient index, 3.10 for CK vs. 2.85 and 2.61 for GK1 and GK2, respectively (p < 0.001).
Conclusions: Both CK and GK systems appear to provide a comparably high quality dosimetric planning for perioptic tumors, although slight differences in dosimetric parameters between these modalities exist.
- Author(s)
- 곽정원; 변유경; 서원섭; 송시열; 안승도; 윤경준; 이도희; 이상욱; 조병철; 조영현
- Issued Date
- 2021
- Type
- Article
- URI
- https://oak.ulsan.ac.kr/handle/2021.oak/8307
- Publisher
- International Journal of Radiosurgery
- Location
- 대한민국
- Language
- 영어
- Citation Volume
- 10
- Citation Number
- 1
- Citation Start Page
- 28
- Citation End Page
- 35
-
Appears in Collections:
- Medicine > Medicine
- 공개 및 라이선스
-
- 파일 목록
-
Items in Repository are protected by copyright, with all rights reserved, unless otherwise indicated.