Technical Feasibility and Histological Analysis of Balloon-Expandable Metallic Stent Placement in a Porcine Eustachian Tube

Metadata Downloads
Abstract
There is a clinical need to develop a stent to treat obstructive and refractory Eustachian tube dysfunction (ETD) after balloon Eustachian tuboplasty. An animal model for stent placement in the Eustachian tube (ET) is needed to develop optimal designs and materials, as stents for ETD have not been clinically applied. The purpose of this study was to evaluate the technical feasibility of stent placement and histological changes in a porcine ET model. Six ETs were evaluated in three pigs. Cobalt-chrome alloy stents with two different diameters were placed in the left and right ET of each animal (right, 3.5 mm; left, 2.5 mm). The outcomes were assessed by endoscopic and fluoroscopic imaging during the procedure, computed tomography after the procedure, and by histological examinations. Stent placement was technically successful in all specimens after metallic guiding sheaths were located in the nasopharyngeal end of the ET. The mean luminal diameters of the proximal, middle, and distal portions of the larger stents in the right ETs were 3.48 mm, 2.54 mm, and 2.15 mm, respectively. In the left ETs using smaller stents, these values were 2.49 mm, 1.73 mm, and 1.42 mm, respectively. The diameters of the inserted stents differed by stent location and the original diameter. Histological findings showed tissue hyperplasia with severe inflammatory cell infiltration at 4 weeks after stent placement. In conclusion, stent placement into the porcine ET was technically feasible, and stent-induced tissue hyperplasia was significantly evident. The luminal configuration of the placed ET stent changed according to its non-elastic nature and anatomical features of the porcine ET. Using this model, ET stents of various materials and designs with anti-inflammatory or anti-proliferative drugs can be optimized for future treatments of ET dysfunction.
Author(s)
강우석강정민곽민영김예리류대성박정훈박홍주송호영
Issued Date
2021
Type
Article
Keyword
Eustachian tubeEustachian tube dysfunctionotitis mediastenttissue hyperplasia
DOI
10.3390/app11041359
URI
https://oak.ulsan.ac.kr/handle/2021.oak/8521
https://ulsan-primo.hosted.exlibrisgroup.com/primo-explore/fulldisplay?docid=TN_cdi_doaj_primary_oai_doaj_org_article_0c1b01e213944fb1838bf49690091388&context=PC&vid=ULSAN&lang=ko_KR&search_scope=default_scope&adaptor=primo_central_multiple_fe&tab=default_tab&query=any,contains,Technical%20Feasibility%20and%20Histological%20Analysis%20of%20Balloon-Expandable%20Metallic%20Stent%20Placement%20in%20a%20Porcine%20Eustachian%20Tube&offset=0&pcAvailability=true
Publisher
APPLIED SCIENCES-BASEL
Location
스위스
Language
한국어
ISSN
2076-3417
Citation Volume
11
Citation Number
4
Citation Start Page
0
Citation End Page
0
Appears in Collections:
Medicine > Medicine
공개 및 라이선스
  • 공개 구분공개
파일 목록
  • 관련 파일이 존재하지 않습니다.

Items in Repository are protected by copyright, with all rights reserved, unless otherwise indicated.