KLI

Standardization Status of Total Cholesterol Concentration Measurement: Analysis of Korean External Quality Assessment Data

Metadata Downloads
Abstract
Background: Total cholesterol concentration measurement is important in the diagnosis of dyslipidemia and evaluation of cardiovascular disease risk factors. Measurement reliability for obtaining an accurate total cholesterol concentration requires procedure standardization. We evaluated the standardization status for total cholesterol concentration measurement through Korean external quality assessment (EQA) data analysis.

Methods: This study involved 1,670 laboratories that participated in the EQA of total cholesterol concentration measurements in 2019 for 32 products from different manufacturers. The target concentrations of three quality control (QC) materials (samples A, B, and C) were measured using the reference method and compared with EQA data. The performance criteria for total cholesterol concentration measurement were based on the National Cholesterol Education Program guidelines, with +/- 3% inaccuracy.

Results: The target values and inaccuracies of the QC material based on the reference method measurements were 254.65 +/- 7.64, 108.30 +/- 3.25, and 256.29 +/- 7.69 mg/dL (6.59 +/- 0.20, 2.80 +/- 0.08, and 6.63 +/- 0.20 mmol/L) for samples A, B, and C, respectively. The performance criteria were not met in 42.7% laboratories for sample A, 68.4% of laboratories for sample B, and 38.0% laboratories for sample C.

Conclusions: Despite significant efforts to accurately measure total cholesterol concentrations, further actions are needed for measurement standardization. Manufacturers reporting values that differ from target values should check calibrator traceability; additional efforts to accurately measure total cholesterol concentrations are required for laboratories that use products from these manufacturers.
Author(s)
윤영안이용화김솔잎이경훈박형두전사일민원기
Issued Date
2021
Type
Article
Keyword
InaccuracyTotal cholesterolExternal quality assessmentStandardization
DOI
10.3343/alm.2021.41.4.366
URI
https://oak.ulsan.ac.kr/handle/2021.oak/8583
https://ulsan-primo.hosted.exlibrisgroup.com/primo-explore/fulldisplay?docid=TN_cdi_nrf_kci_oai_kci_go_kr_ARTI_9798940&context=PC&vid=ULSAN&lang=ko_KR&search_scope=default_scope&adaptor=primo_central_multiple_fe&tab=default_tab&query=any,contains,Standardization%20Status%20of%20Total%20Cholesterol%20Concentration%20Measurement:%20Analysis%20of%20Korean%20External%20Quality%20Assessment%20Data&offset=0&pcAvailability=true
Publisher
ANNALS OF LABORATORY MEDICINE
Location
대한민국
Language
영어
ISSN
2234-3806
Citation Volume
41
Citation Number
4
Citation Start Page
366
Citation End Page
371